Draeger X-Pid® 9500

Draeger X-Pid® 9500

Draeger X-Pid® 8500

SKU
X-PID

Draeger X-pid® 8500 -

The Draeger X-pid® 8500 is a selective PID gas measurement device that is idealfor users who frequently test for hazardous toxic substances. Benzene, butadiene and other volatile organic compounds (VOCs) are cacinogenic even in the smallest concentrations. Selective measurement is necessary because other gases and vapors are often also present. The Draeger X-pid® 8500 allows for short test times and laboratory-quality results.

The Draeger X-pid® 8500 has pre-defined target compunds that can be accurately measured in seconds. The Draeger X-pid® is a streamlined version of gas chromatography analyses conducted in the laboratory. But it builds on that as the user now can easily take the analysis to the field where intrinsic safety certifications may be required. 

The wireless (bluetooth) handheld Class 1 Div 1 Control Unit with built in mobil app makes using the sensor unit lightweight and very easy to use. The Draeger X-pid® 8500 is optimized for measuring in this concentration range and detects benzene concentrations as low as 50ppb.

Draeger X-pid® 8500 Literature

Draeger X-pid® 8500 
Target Compunds, LOQ and upper range.
 

Target Compunds CAS Number Retention time, s LOD1, ppm LOQ2, ppm UR3, ppm
Acetone 67-64-1 8.1 0.17 0.50 50.0
Acrolein  107-02-8  7.8  0.33  1.00  100
Benzene  71-43-2 19.3 0.02 0.05 25.0
Butadiene, 1,3-  106-99-0 6.4 0.07  0.20 25.0
Butanone, 2-  78-93-3 12.9 1.00 3.00 300
Butyl acrylate  141-32-2 125.5 - - -
Chlorobenzene  108-90-7 15.6 1.00 3.00 200
Dichloroethene, 1,1-  75-35-4 8.9 0.07 0.20 50.0
Dichloroethene, cis-1,2-  156-59-2 13.4 0.07 0.20 50.0
Dichloroethene, trans-1,2-  156-60-5 10.9 0.07 350 50.0
Epichlorohydrin  106-89-8 27.3 0.67 2.00 200
Ethyl acetate  141-78-6 14.6 1.00 3.00 300
Ethyl acrylate  140-88-5 24.9 1.00 3.00 200
Ethylbenzene  100-41-4 88.7 1.00 3.00 300
Ethylene oxide  75-21-8 6.8 0.33 1.00 100
Hexane, n-  110-54-3  13.7 0.33 1.00 100
Isobutylene  115-11-7 6.3 0.07 0.20 100
Methyl acrylate  96-33-33 14.4 0.67 2.00 200
Methyl bromide  74-83-9 6.8 0.17 0.50 100
Phosphine 7803-51-2 5.3 0.67 2.00 100
Propylene oxide  75-56-9 8.2 0.17 0.50 25.0
Styrene  100-42-5 111.3 1.00 3.00 300
Tetrachloroethylene  127-18-4 58.9 0.67 2.00 150
Tetrahydrofuran  109-99-9  16.5  1.00  3.00  200
Toluene  108-88-3  41.6  0.33  1.00  100
Trichloroethylene  79-01-6  24.9  0.33  1.00  100
Vinyl chloride  75-01-4  6.3  0.33  1.00  100
Xylene, m-  108-38-3  95.7  1.00  3.00  300
Xylene, o-  95-47-6  114.5  1.00  3.00  300
Xylene, p-  106-42-3  96.6 1.00 3.00 300

1 Limit of Detection; 2 Limit of Quantification 

> 150 additional compunds can be qualified.

The technical specifications refer exclusively to the Dräger X-pid® 8500 sensor unit.

Specifications
Dimensions: (W x H x D) Approx. 5.2 x 11.1 x 2.2 in
Weight Approx.: 880 g (2 lb)
Temperature: 14 °F to 95 °F (-10 to +35 °C)
Ambient conditions
Pressure:
700 to 1,300 mbar 
Ambient conditions
Relative humidity:
10% to 95% RH
Protection class: IP54
Start-up phase: Approx. 10 min *can be increased at low ambient temperatures
Operating times: Typically 6 h, reduces with lower ambient temperatures
Approvals Sensor Unit:

Ex-Zone:
ATEX:
IECEx:
cCSAus:


0

II 1G Ex ia IIC T4 Ga IP54
Ex ia IIC T4 Ga IP54
Class I, Div 1 Groups A-D T4 
Approvals Control Unit:
(ecom Smart-Ex 01)

Ex-Zone:
ATEX:
IECEx:
cCSAus:



1
II 2G Ex ib IIC T4 Gb IP64
Ex ib IIC T4 Gb IP64
Class I, Div 1 Groups A-G T4


Additional X-pid® Specifications:
X-pid® specifications

Standard X-pid System Includes:

- X-pid® Sensing Unit with charger / power supply
- X-pid® Control Unit with pre-installed mobil app and charger / power supply
- Calibration Tube and adapter
- 2ea Sensing Unit Filters
- 3ea Luer Connectors
- Shoulder Strap
- Control Unit Belt Clip / Holder
- Manual and Supporting Calibration / Test Information

Accessory/Calibration Kit #1 Includes/adds:
- Custom Cut Draeger Plastic Case that will also hold calibration equipment or call us Pelican Cases available.
- 1ea Fixed Flow Regulator 
- 1ea 116 Liter Calibration Gas Isobutylene / Toluene Mix

Accessory/Calibration Kit #2 Includes/adds:
- Custom Cut Draeger Plastic Case that will also hold calibration equipment or call us Pelican Cases available.
(only one cylinder fits in Draeger Case)
- 2ea Fixed Flow Regulator 
- 1ea additional Calibration Tube and adapter
- 1ea 116 Liter Isobutylene / Toluene Mix Calibration Gas 
- 1ea 116 Liter Benzene Calibration Gas

** Sensing Unit and Control Unit can be purchased seperately. You must have a control unit to operate the Sensing Unit! 

Cost savings -
Because no consumables are used to take measurements, operating costs can be reduced. For users with high measurement needs, the Dräger X-pid 8500 quickly pays for itself. For example, it requires no pre-tubes, also making it easier to use and helps prevent user errors. On the basis of about 200 measurements per year, the Dräger X-pid 8500 is generally more cost-effective than comparable yet more cumbersome measurement systems.

Time savings of up to 90% -
The Dräger X-pid 8500 requires no preparation and is ready to use after a brief start-up phase. Selective measurement takes only a few seconds. A benzene measurement starts with the push of a button and is completed in only 30 seconds. After another 60 seconds, the device is ready to measure for benzene again. Compared with other detection systems, the Dräger X-pid 8500 saves considerable time and enables further monitoring to be undertaken. Simultaneous measurement of other compounds further reduces testing times.

Reliable performance under tough conditions -
The influence on measurement results by environmental factors like variable ambient temperature or high humidity are reduced to a minimum. The sensor unit maintains a constant temperature above the ambient air temperature and separates water vapor from the target compounds. This ensures reliable measurements under tough environmental conditions.

Simple functional test and easy calibration -
The functional test with isobutylene and toluene test gases is done in just two minutes and the Dräger X-pid 8500 is ready to use. During the functional test the user is guided through the user interface step-bystep. Only if the functional test fails, a calibration with the same test gas standard will be required, which will be completed in about four minutes.